APOLLO 3-PORT OPERATING INSTRUCTIONS

apollo 3 port alternating pressure mattress operating manual  

Setting up the System

  1. - Un-pack the mattress and place on top of the hospital bed and attach the adjustable straps around the four corners of the hospital bed frame.
  2. Make sure that the CPR release valve is at the left head side of the bed and is clear of the bed rails.
  3. - Attach hose from the mattress to the input located on side of pump. Push connector till it locks in place.
  4. Secure the pump to the footboard of the bed with the fold-away bed hanger brackets  (LOCATED ON THE BACK OF THE PUMP.)
  5. Plug the pump into a grounded electrical outlet.
  6. Turn pump to the on position and adjust dial to maximum output
  7. To fill the mattress quickly, use the optional orange inflator. Insert the inflator into the CPR PORT located on the left head side of the bed. Twist the CPR to the open position and fill till all bladders are firm.  Once filled, twist the CPR PORT to the closed position. ALLOW SYSTEM TO RUN ONE COMPLETE 10-MINUTE CYCLE.

  

Pump Setting Chart 

PATIENT WEIGHT LBS.

SETTING

85 - 140

1 - 2

120 - 165

1 - 3

150 - 195

1 - 4

175 - 205

2 - 4

200 - 250

3 - 6

245 - 300

4 - 6

300 - 350

7 - 8

 

Operating Procedures

  • Once mattress is fully inflated, place patient on to the mattress. Allow the system to run a full cycle (10 min.).
  • Adjust the comfort control knob to the patient’s desired level.
  • A hand check can be performed by placing your hand under the patient on a deflated bladder. Make sure the patient is not touching the base of the bed. If this occurs      increase the output by turning the control dial higher.
  • In the event of cardiac arrest, twist the  CPR Valve to the open position to release air in the mid torso section. It will take less than 10 seconds to deflate.

 

Cleaning and Laundering Procedures

  • Disconnect power cord from electrical outlet.
  • The mattress can be wiped down between patients using hospital detergent or a mixture of two ounces of bleach per one gallon of water
  • For laundering remove cover by unzipping both sides. Wash for at least 15-20 minutes in the gentle cycle.
  • Tumble dry at very low or no heat
  • Individual bladders can be washed as well but must be capped with part #8E0010 to restrict water entering the bladder.

Disassembling the System

  • Unplug Pump from electrical outlet.
  • Disconnect mattress hoses from pump. 
  • Turn Handle on CPR Valve to release air.
  • Insert orange deflator into the CPR Valve and evacuate air until mattress is totally flat.
  • Tuck hoses in mattress and roll it to fit into carry bag.

Safety Tips

  • Do not expose the pump or power cord to liquids of any kind
  • Always protect the system from open flames
  • Avoid the use of hypercarbonate or phenol based solutions in cleaning
  • Avoid the use of sharp objects near the mattress
  • Always store the mattress clean and dry
  • Consider the appropriate use of side rails on the patients bed